User`s guide
Agilent B1500 VXIplug&play Driver User’s Guide, Edition 3 2-21
Driver Functions
agb1500_cmdInt16Arr_Q
value  Parameter for command. -2147483647 to 2147483647.
agb1500_cmdInt16Arr_Q
This function passes the cmd_str string to the instrument. This function expects a 
response that is a definite arbitrary block of 16 bit integers. You specify the cmd_str 
and size parameters, and get result[ ] and count.
Syntax ViStatus _VI_FUNC agb1500_cmdInt16Arr_Q(ViSession vi, ViString cmd_str, 
ViInt32 size, ViInt16 _VI_FAR result[ ], ViPInt32 count);
Parameters vi  Instrument handle returned from agb1500_init( ).
cmd_str  Instrument command (cannot exceed 256 bytes in length).
size  Size of result[ ] (number of items in the array). 
1 to 2147483647.
result[ ]  Response from instrument.
count  Count of valid items in result[ ]. Returned data.
agb1500_cmdInt16_Q
This function passes the cmd_str string to the instrument. This function expects a 
response that can be returned as a 16 bit integer.
Syntax ViStatus _VI_FUNC agb1500_cmdInt16_Q(ViSession vi, ViString cmd_str, 
ViPInt16 result);
Parameters vi  Instrument handle returned from agb1500_init( ).
cmd_str  Instrument command (cannot exceed 256 bytes in length).
result  Response from instrument.
agb1500_cmdInt32Arr_Q
This function passes the cmd_str string to the instrument. This function expects a 
response that is a definite arbitrary block of 32 bit integers. You specify the cmd_str 
and size parameters, and get result[ ] and count.
Syntax ViStatus _VI_FUNC agb1500_cmdInt32Arr_Q(ViSession vi, ViString cmd_str, 
ViInt32 size, ViInt32 _VI_FAR result[ ], ViPInt32 count);