User`s guide
3- 46 Keysight B1500A User’s Guide, Edition 14
Installation
Connecting Measurement Devices
Using Connector Plate
This section provides the information useful for connecting cables and probing 
needles to a connector plate.
• “To Reduce Leakage Current”
• “To Measure Low Resistance”
To Reduce Leakage Current
To reduce the leakage current caused by connection cables, the guard technique is 
effective. Connect the probing needles to the terminals of the connector plate by 
using coaxial cables as shown below:
1. At end of cable, connect coaxial center conductor to force terminal of connector 
plate, and connect coaxial outer conductor to guard terminal of connector plate.
2. At another end, connect coaxial center conductor to tail of the probing needle. 
Never connect the outer conductor at this cable end.
Extend the outer conductor as close as possible to the probing needle.
Example 
The following example connection can be used to reduce the leakage current. 
Extend the outer conductor as close as possible to the probing needle. This also 
reduces the induced noise.
WARNING For safety, prepare and attach a shield cover to the connector plate to avoid 
touching the joints on the wiring side of connectors. And ground the cover. 
Wrap an insulator around the naked conductors of coaxial cable to avoid 
touching it.
Probing needle 
Contact pad 
Connector plate 
Outer conductor
Center conductor
Coaxial cable 