User`s guide
4- 24 Keysight B1500A User’s Guide, Edition 14
Specifications
Hardware
• Measurement accuracy
Measurement accuracy at the measurement port of the MFCMU or the MFCMU 
cable (1.5 m or 3.0 m)
• |Z| accuracy: ± E (%)
•  accuracy: ± E/100 (radian)
• C accuracy:
± E (%) at Dx  0.1
± E(1+Dx
2
)
1/2
 (%) at Dx > 0.1
• D accuracy:
± E/100 at Dx  0.1
± E(1+Dx)/100 at Dx > 0.1
• G accuracy:
± E/Dx (%) at Dx  0.1
± E(1+Dx
2
)
1/2
/Dx (%) at Dx > 0.1
where,
Zx: Impedance measurement value ()
Dx: Measurement value of D
E = E
P
’ + (Z
S
’ / |Zx| + Y
O
’ |Zx| )  100 (%)
E
P
’ = E
PL
 + E
POSC
 + E
P
 (%)
Y
O
’ = Y
OL
 + Y
OSC
 + Y
O
 (S)
Z
S
’ = Z
SL
 + Z
OSC
 + Z
S
 () 
For the parameter value, see Table 4-11 to Table 4-13. Calculation example of 
the measurement accuracy is also shown in Table 4-14.
Specification conditions:
• Temperature: 23 C ± 5 C
• Integration time: 1 PLC or 16 PLC