User`s guide
4- 40 Keysight B1500A User’s Guide, Edition 14
Specifications
Hardware
Table 4-22 Minimum timing parameters for current measurement
1
Voltage applied to DUT 10 V
Current applied to DUT 100 nA 1 A10 A 100 A 1 mA 10 mA
Applied 
voltage 
condition
Recommended 
minimum pulse 
width
a
a. Recommended minimum pulse width = settling time + recommended minimum measurement 
window.
47 s38.7 s6.8 s 950 ns 240 ns 145 ns
Current 
measurement 
condition
Measurement Range 1 A1 A10 A 100 A 1 mA 10 mA
Recommended 
minimum 
measurement 
window
10 s 1.64 s1 s 130 ns 40 ns 20 ns
Settling time
b
b. The time until the measured value settles to within ± 0.6 % of the final result value after the out-
put voltage is changed from the initial value (0 V). Minimum rise/fall time of 70 ns is recom-
mended for minimizing overshoot. 
37 s37 s5.8 s 820 ns 200 ns 125 ns
Noise (rms)
c
c. RMS noise measured over the recommended minimum measurement window. 
160 pA 425 pA 2.5 nA 47 nA 280 nA 1.9 A
1. Measurement conditions: The DUT is a resistive load chosen to adjust the flowing current to the 
specified current in the table above. The capacitance of the cable between the RSU and the DUT 
is 20 pF. Voltage is applied to the DUT by a channel of WGFMU/RSU in Fast IV mode and in 
the 10 mA range, and current measurement is performed by another channel at 0 V in Fast IV 
mode.