Installation
WARNING There are potentially hazardous voltages (± 200 V for HPSMU, and ± 100 V for 
MPSMU/HRSMU) present at the Force, Sense, and Guard terminals of this 
instrument. To prevent electrical shock, the following safety precautions must 
be observed during the use of the instrument.
• Use a three-conductor AC power cable to connect the instrument to an 
electrical ground (safety ground).
•If you do not use the B1500A-A5F/16442A/B test fixture, you must install 
and connect an interlock circuit that opens the Interlock terminal when the 
shielding box access door is opened.
• If you change the DUT interface, test fixture, prober, and such, connect an 
interlock cable to the one actually used.
• Confirm periodically that the interlock function works normally.
• Before touching the connections on the Force, Guard, and Sense terminals 
in the test fixture or the shielding box, turn the instrument off and 
discharge any capacitors. If you do not turn the instrument off, complete all 
of the following items, regardless of the instrument settings.
• Press the front panel 
Stop key to set the module output off. 
• Confirm that the front panel High Voltage indicator is not lit.
• Open the Interlock terminal (open the fixture cover or the shielding box 
access door).
• Discharge any capacitors connected to a measurement resource.
• Warn persons working around this instrument about dangerous conditions.