Technical data
Agilent B2900 SCPI Command Reference, Edition 2 4- 59
Subsystem Commands
:OUTPut:RECall
Syntax  :OUTPut[c]:PROTection[:STATe] mode
:OUTPut[c]:PROTection[:STATe]?
Parameter  mode  0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables the over voltage/current protection.
mode=1 or ON enables the over voltage/current protection.
Query response  mode <newline>
mode is 0 or 1, and indicates that the over voltage/current protection is off or on, 
respectively. Response data type is NR1.
Example  :OUTP:PROT 0
:OUTP2:PROT:STAT?
:OUTPut:RECall
Recalls the channel setup saved by the :OUTPut:SAVE command.
Syntax  :OUTPut[c]:RECall index
Parameter  index  0|1. Parameter data type is NR1.
index=0 is used to recall channel setup 0. 
index=1 is used to recall channel setup 1.
Example  :OUTP:REC 1
:OUTPut:SAVE
Saves the channel setup. The setup can be recalled by the :OUTPut:RECall 
command.
Syntax  :OUTPut[c]:SAVE index
Parameter  index  0|1. Parameter data type is NR1.
index=0 is used to memorize the present channel setup as channel setup 0.
index=1 is used to memorize the present channel setup as channel setup 1.