Technical data
Agilent B2900 SCPI Command Reference, Edition 2 4- 73
Subsystem Commands
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:APERture
SENSe Subsystem
For the numeric suffix [c], see “Numeric Suffix” on page 1-8.
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:
APERture
Sets the integration time for one point measurement.
Syntax  :SENSe[c]:<CURRent[:DC]|RESistance|VOLTage[:DC]>:APERture time
:SENSe[c]:<CURRent[:DC]|RESistance|VOLTage[:DC]>:APERture? [time]
For <CURRent[:DC]|RESistance|VOLTage[:DC]>, set CURRent[:DC], 
RESistance, or VOLTage[:DC]. Specifying the measurement item is not important 
because the time value is common for all items.
Parameter  time  value (+8E-6 to +2 seconds)|MINimum|MAXimum|DEFault 
(default is 0.1 PLC, =0.1/power line frequency). Parameter data 
type is NRf+. Query does not support time=value. If you 
specify the value less than MIN or greater than MAX, time is 
automatically set to MIN or MAX.
The integration time can be expressed by the following formula, using the NPLC 
value set by the 
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:NPLCycles 
command. So the last command setting is effective for both time and nplc.
time = nplc / power line frequency
Query response  time <newline>
time returns the present setting of the integration time. If a parameter is specified, 
time returns the value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example  :SENS:CUR:APER 2E-3
:SENS2:CURR:DC:APER?
:SENSe:<CURRent[:DC]|RESistance|VOLTage[:DC]>:
APERture:AUTO
Enables or disables the automatic aperture function.