Technical data
4- 90 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
:SENSe:WAIT:GAIN
:SENSe:WAIT:GAIN
Sets the gain value used for calculating the measurement wait time for the specified 
channel. See 
:SENSe:WAIT[:STATe].
Syntax  :SENSe[c]:WAIT:GAIN gain
:SENSe[c]:WAIT:GAIN? [gain]
Parameter  gain  value (0 to 100)|MINimum|MAXimum|DEFault (default is 1). 
Parameter data type is NRf. Query does not support gain=value.
Query response  gain <newline>
gain returns the present setting of the gain value. If a parameter is specified, gain 
returns the value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example  :SENS:WAIT:GAIN 0.5
:SENS2:WAIT:GAIN?
:SENSe:WAIT:OFFSet
Sets the offset value used for calculating the measurement wait time for the 
specified channel. See 
:SENSe:WAIT[:STATe].
Syntax  :SENSe[c]:WAIT:OFFSet offset
:SENSe[c]:WAIT:OFFSet? [offset]
Parameter  offset  value (0 to 1 seconds)|MINimum|MAXimum|DEFault (default 
is 0). Parameter data type is NRf. Query does not support 
offset=value.
Query response  offset <newline>
offset returns the present setting of the offset value. If a parameter is specified, offset 
returns the value assigned to DEF, MIN, or MAX. Response data type is NR3.
Example  :SENS:WAIT:OFFS 0.5
:SENS2:WAIT:OFFS?