Technical data
4- 96 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Commands
[:SOURce]:<CURRent|VOLTage>:RANGe:AUTO
Example  :CURR:RANG 1E-6
:SOUR2:VOLT:RANG?
[:SOURce]:<CURRent|VOLTage>:RANGe:AUTO
Enables or disables the automatic ranging function for the specified source channel.
Syntax  [:SOURce[c]]:<CURRent|VOLTage>:RANGe:AUTO mode
[:SOURce[c]]:<CURRent|VOLTage>:RANGe:AUTO?
For <CURRent|VOLTage>, specify CURRent for current output, or VOLTage for 
voltage output.
Parameter  mode  0|OFF|1|ON (default). Parameter data type is boolean.
mode=0 or OFF disables automatic ranging. If this function is disabled, the source 
output is performed by using the range set by the 
[:SOURce]:<CURRent|VOLTage>:RANGe command.
mode=1 or ON enables automatic ranging. If this function is enabled, the channel 
automatically sets the range which provides the best resolution to apply the source 
output.
If a range is manually selected, automatic ranging is disabled.
Query response  mode <newline>
mode is 0 or 1, and indicates that automatic ranging is off or on, respectively. 
Response data type is NR1.
Example  :CURR:RANG:AUTO 0
:SOUR2:VOLT:RANG:AUTO?
[:SOURce]:<CURRent|VOLTage>:RANGe:AUTO:LLI
Mit
Specifies the lower limit for the automatic output ranging operation, and sets the 
minimum range which provides the best resolution to apply the specified value.
Syntax  [:SOURce[c]]:<CURRent|VOLTage>:RANGe:AUTO:LLIMit range
[:SOURce[c]]:<CURRent|VOLTage>:RANGe:AUTO:LLIMit? [range]