:ABORt<:ACQuire|:TRANsient|[:ALL]>
TRIGger Subsystem
For the numeric suffix [c], see “Numeric Suffix” on page 1-8.
:ABORt<:ACQuire|:TRANsient|[:ALL]>
Aborts the specified device action for the specified channel. Trigger status is 
changed to idle. 
Syntax  :ABORt<:ACQuire|:TRANsient|[:ALL]> [chanlist]
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement, 
:TRANsient for source output, or :ALL for both device actions.
Parameter  chanlist  Channels. Parameter data type is channel list. 
(@1)|(@2)|(@1,2)|(@1:2)|(@2,1)|(@2:1). See 
“Channel List 
Parameter” on page 1-8.
(@1) selects channel 1 only. (@2) selects channel 2 only. (@1,2), (@1:2), (@2,1), 
and (@2:1) selects both channels 1 and 2.
If this parameter is not specified, chanlist=(@1) is set.
Example  :ABOR:ACQ (@2)
:ARM<:ACQuire|:TRANsient|[:ALL]>[:IMMediate]
Sends an immediate arm trigger for the specified device action to the specified 
channel. When the status of the specified device action is initiated, the arm trigger 
causes a layer change from arm to trigger. 
Syntax  :ARM<:ACQuire|:TRANsient|[:ALL]>[:IMMediate] [chanlist]
For <:ACQuire|:TRANsient|[:ALL]>, specify :ACQuire for measurement, 
:TRANsient for source output, or :ALL for both device actions.
Parameter  chanlist  Channels. Parameter data type is channel list. 
(@1)|(@2)|(@1,2)|(@1:2)|(@2,1)|(@2:1). See 
“Channel List 
Parameter” on page 1-8.
(@1) selects channel 1 only. (@2) selects channel 2 only. (@1,2), (@1:2), (@2,1), 
and (@2:1) selects both channels 1 and 2.