Technical data
5- 4 Agilent B2900 SCPI Command Reference, Edition 2
Error Messages
Command Error
Command Error 
If syntax of SCPI command is not valid, a -1XX error occurs.
Error -100  Command error
Generic syntax error that cannot be determined more specifically.
Error -101  Invalid character
An invalid character for the type of a syntax element was received; for example, a 
header containing an ampersand.
Error -102  Syntax error
An unrecognized command or data type was received; for example, a string was 
received when B2900 does not accept strings.
Error -103  Invalid separator
An illegal character was received when a separator was expected; for example, the 
semicolon was omitted after a program message unit.
Error -104  Data type error
An improper data type was received; for example, numeric data was expected but 
string data was received.
Error -105  GET not allowed
A group execute trigger was received within a program message.
Error -108  Parameter not allowed
Too many parameters for the command were received.
Error -109  Missing parameter
Fewer parameters were received than required for the command.
Error -110  Command header error
An error was detected in the header. This error message is reported if B2900 cannot 
determine the more specific header errors -111 through -114.