Technical data
Agilent B2900 SCPI Command Reference, Edition 2 2- 7
Subsystem Command Summary
Setting Source/Measure Unit
[:SOUR[c]]:CURR:MODE mode
[:SOUR[c]]:VOLT:MODE mode
[:SOUR[c]]:CURR:MODE?
[:SOUR[c]]:VOLT:MODE?
Selects the source mode, fixed, list 
sweep, or sweep, of the specified source 
channel.
mode=SWEep|LIST|FIXed
FIX
[:SOUR[c]]:CURR:POIN points
[:SOUR[c]]:VOLT:POIN points
[:SOUR[c]]:CURR:POIN? [DEFault|M
INimum|MAXimum]
[:SOUR[c]]:VOLT:POIN? [DEFault|M
INimum|MAXimum]
Sets the number of sweep steps for the 
current or voltage sweep output.
points=MINimum|MAXimum|DEFault| 
1
 to 2500
The points value can be expressed by the 
following formula, using the step value 
set by the 
[:SOUR[c]]:<CURR|VOLT>:STEP 
command and the span value set by the 
[:SOUR[c]]:<CURR|VOLT>:SPAN 
command. 
points = span/step + 1 (where step is not 
0)
points=1 sets step=0.
1
[:SOUR[c]]:CURR:RANG range
[:SOUR[c]]:VOLT:RANG range
[:SOUR[c]]:CURR:RANG?
[:SOUR[c]]:VOLT:RANG?
Sets the current or voltage output range of 
the specified source channel. This 
command is effective when the automatic 
ranging function is off.
range=MINimum|MAXimum|DEFault| 
minimum to maximum source value, in A 
or V. See 
“Source Output Ranges” on 
page 2-24.
1.00E-04 A 
for current 
range, or 2 V 
for voltage 
range
[:SOUR[c]]:CURR:RANG:AUTO mode
[:SOUR[c]]:VOLT:RANG:AUTO mode
[:SOUR[c]]:CURR:RANG:AUTO?
[:SOUR[c]]:VOLT:RANG:AUTO?
Enables or disables the automatic ranging 
function for the specified source channel.
mode=1|ON|0|OFF
ON
Command Summary Reset setting