Technical data
2- 16 Agilent B2900 SCPI Command Reference, Edition 2
Subsystem Command Summary
Setting Source/Measure Unit
:SENS[c]:CURR[:DC]:NPLC:AUTO mod
e
:SENS[c]:RES:NPLC:AUTO mode
:SENS[c]:VOLT[:DC]:NPLC:AUTO mod
e
:SENS[c]:CURR[:DC]:NPLC:AUTO?
:SENS[c]:RES:NPLC:AUTO?
:SENS[c]:VOLT[:DC]:NPLC:AUTO?
Enables or disables the automatic NPLC 
function.
mode=1|ON|0|OFF
The automatic NPLC on/off works with 
the automatic aperture on/off set by the 
:SENS:<CURR|RES|VOLT>:APER:AU
TO command. So the last command 
setting is effective for both functions.
OFF
:SENS[c]:CURR[:DC]:PROT[:LEV] co
mp
:SENS[c]:VOLT[:DC]:PROT[:LEV] co
mp
:SENS[c]:CURR[:DC]:PROT[:LEV]? [
DEFault|MINimum|MAXimum]
:SENS[c]:VOLT[:DC]:PROT[:LEV]? [
DEFault|MINimum|MAXimum]
Sets the compliance value of the specified 
channel.
comp=MINimum|MAXimum|DEFault| 
minimum to maximum measurement 
value, in A or V. See 
“Measurement 
Ranges” on page 2-27.
1.00E-04 A 
for current 
compliance, 
2.0 V for 
voltage 
compliance
:SENS[c]:CURR[:DC]:PROT:TRIP?
:SENS[c]:VOLT[:DC]:PROT:TRIP?
Returns the compliance status of the 
specified channel
Response is 1 or 0 that indicates the 
channel is in the compliance state or not.
:SENS[c]:CURR[:DC]:RANG:AUTO mod
e
:SENS[c]:RES:RANG:AUTO mode
:SENS[c]:VOLT[:DC]:RANG:AUTO mod
e
:SENS[c]:CURR[:DC]:RANG:AUTO?
:SENS[c]:RES:RANG:AUTO?
:SENS[c]:VOLT[:DC]:RANG:AUTO?
Enables or disables the automatic ranging 
function of the specified measurement 
channel.
mode=1|ON|0|OFF
ON
Command Summary Reset setting