Technical data
Agilent B2900 SCPI Command Reference, Edition 2 2- 27
Subsystem Command Summary
Setting Source/Measure Unit
Measurement Ranges
Table 2-7 Voltage Measurement Range
Table 2-8 Current Measurement Range
Range value Voltage measurement value Resolution
0.2 V 0  |V|  0.212 V 0.1 V
2 V 0  |V|  2.12 V 1 V
20 V 0  |V|  21.2 V 10 V
200 V 0  |V|  212 V 100 V
Range value Current measurement value Resolution
10 nA 
a
a. Available for the B2911A and B2912A. Not available for the B2901A 
and B2902A.
0  |I|  10.6 nA 10 fA
100 nA 0  |I|  106 nA 100 fA
1 A 0  |I|  1.06 A 1 pA
10 A 0  |I|  10.6 A 10 pA
100 A 0  |I|  106 A 100 pA
1 mA 0  |I|  1.06 mA 1 nA
10 mA 0  |I|  10.6 mA 10 nA
100 mA 0  |I|  106 mA 100 nA
1 A 0  |I|  1.06 A 1 A
1.5 A 0  |I|  1.53 A
3 A 0  |I|  3.06 A 10 A
10 A 
b
b. Available for pulse mode. Not available for DC mode.
0  |I|  10.6 A