Specifications

58
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen  ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪  info@datatec.de ▪ www.datatec.de
16008B Resistivity Cell
Terminal Connector: 
* InterlockConnectorenablesanddisablestheapplicationo
source voltage from the measurement instrument.
DUT Connection: 2-Terminal (with triaxial cable)
Dimensions (approx.): 240(W) x 180(H) x 240(D) [mm]
Cable Length (approx.): 1.2 m (connector to electrodes)
Weight (approx.): 7000 g
58
DC (High resistance)  Material
Description:  The  16008B  is  designed  to  operate  specifically  with 
4339B. It is used to measure surface or volume resistance/resistivi-
ty of insulation materials. The following figures show the block dia-
grams of resistivity measurements. 
Volume  resistivity  is  measured  across  the  material  (MUT)  by  the 
unguarded  and  guarded  electrodes.  Surface  resistivity,  on  the  other 
hand, is measured along the surface of the MUT (between the guard-
ed and  the  unguarded  electrodes). Notice  that  the  role  of  the  guard 
and unguarded  electrodes switch when  measuring volume  and  sur-
face  resistivity. 16008B  is  provided  with  three  different  electrode 
sizes  to  meet  size  requirements  by  different  measurement  stan-
dards. The guarded electrode eliminates measurement errors due to 
the edge effect and arbitrary contact pressure can be applied to the 
material  under  test.  It  is  also  equipped  with  a  high-voltage protec-
tion cover to shut off power when opened.
Applicable Instruments: 4339B*
*denotestheinstrumentisobsolete.
Frequency: DC 
Maximum Voltage: 1000 V
Maximum Current: 10 mA
Resistance Measurement Range: 
Volume Resistivity Measurement Range: 
  up to 4.0 x 10
18
 Ωcm
Surface Resistivity Measurement Range: 
  up to 4.0 x 10
17
 Ω
Operating Temperature: -30°C to +100°C (excluding selector switch) 
Material Size: Select an electrode so that outer diameter of guard-
ing electrode is smaller than the DUT’s diameter. See figure and 
table below for more details:
Volume resistivity measurement  Surface resistivity measurement
Electrode sizes and applicable material sizes
* uterDiameteroGuardElectrode2mm
  Thickness: 10µm to 10 mm
Type Connector
Input Triaxial (special screw-type)
Output High Voltage BNC (special type)
Control Interlockconnector*
D1 D2 D3 Ordering information D
Main Electrode Guard Electrode 
(Inner Diameter)
Guard Electrode 
(Inner Diameter)
— Material Size
26 mm 38 mm 48 mm Supplied with 16008B-001/002 50mm*to125mm
50 mm 70 mm 80 mm Standard - equipped 82mm*to125mm
76 mm 88 mm 98 mm Supplied with 16008B-001 100mm*to125mm