Specifications

71
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen  ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪  info@datatec.de ▪ www.datatec.de
71
Error Compensation
Agilent measurement instruments incorporate one of the following four types of error compensation functions to eliminate 
residual impedance effects in test fixtures:
1. Open/short Compensation
For a simple measurement system, represented by the equivalent circuit model shown below, residual impedance val-
ues Rs and Ls and admittance values Co and Go can be corrected by:
(i)  Measuring open condition for the test fixture’s admittance.
(ii)  Measuring short condition for the test fixture’s impedance
(iii)  Measuring  the  test  sample,  then  subtracting  the  admittance 
and impedance.
The  above  procedure  is  performed  internally  by  the  open/short 
compensation. When  the  test fixture  is directly connected  to the 
measurement  instrument,  the  open/short  compensation  suffi-
ciently corrects the measurement error.
2. Open/short/load Compensation
When  the  measurement  system  is  too  complicated  to  be  repre-
sented  as  the  above  equivalent  circuit  model,  the  open/short  compensation  cannot  completely  compensate  for  the 
residual impedance. In this case, the open/short/load compensation is used rather than the open/short compensation.
The open/short/load compensation is particularly effective when, for example, the 16065A is used with the measure-
ment instrument.
The following table lists measurement instruments and available compensation functions.
–:  N/A
●:  Available
* denotestheinstrumentisobsolete.
Note:  For more details of the compensation functions, refer to the instruction manuals supplied with        
the measurement instrument.
Appendix  Compensation
Test Fixture
Instrument 
DUT
Model Open compensation Short compensation Load compensation Load's input parameters
4263B ● ● ● Z-Q, R-X, Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, 
Cs-Rs, Ls-D, Ls-Q, Ls-Rs
4268A* ● ● ● Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, Cs-Rs
4279A* ● ● ● Cp-D, Cp-G
4284A* ● ● ● same as measurement parameters
4285A ● ● ● same as measurement parameters
4287A* ● ● – –
4288A* ● ● ● Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-Q, Cs-Rs
4291B* ● ● ● Rs-Ls
4294A ● ● ● Rs-Ls
4338B* – ● – –
4339B* ● – – –
4349B* ● – – –
E4980A ● ● ● same as measurement parameters
E4981A ● ● ● Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, Cs-Rs
E4982A ● ● – –
E4991A ● ● – –