4. Cable Length Compensation
When the test sample is measured with an instrument having a Four-Terminal Pair configuration, the additional length 
of the test-lead extension between the instrument and the test sample, in conjunction with the measurement frequen-
cy, influences the amplitude and phase of the signal being measured. This must be taken into account, particularly for 
measurements performed at frequencies of 100 kHz or higher, since the modification of the amplitude and phase of the 
signal can cause the internal measurement circuit to malfunction or create an unexpected measurement error. Such cir-
cuit malfunctions and measurement errors can be avoided with cable length compensation. Cable length compensation 
should therefore  be  performed prior  to the open/short  compensation. When  the  following measurement  instruments 
and  test  leads  are  used  in  combination,  the  undesired  effects  described  above  can  be  eliminated  by  performing  the 
cable length compensation (the values in the table represent cable length settings):
–:  N/A
1
:  Available for 4284A with Option 006.
* denotestheinstrumentisobsolete
Measurement Repeatability
Dirty electrodes on the test fixture cause an increase in the contact resistance. This increase in resistance can cause poor 
measurement repeatability. Therefore,  the electrode  of  the test  fixture must  be kept  clean when  measurements  are per-
formed.
73
Appendix  Compensation
Model 16048A 16048D 16048E 16048G 16048H
4263B 1 m 2 m 4 m – –
4268A* 1 m 2 m – – –
4279A* 1 m 2 m – – –
4284A* 1 m 2 m
1
4 m
1
– –
4285A 1 m 2 m – – –
4288A* 1 m 2 m – – –
4294A – – – 1 m 2 m
E4980A 1 m 2 m 4 m – –
E4981A 1 m 2 m – – –