Technical data

Calibrating the Infiniium Oscilloscope and Probe 14
U7232A DisplayPort Compliance Testing Notes 131
Internal Calibration
This will perform an internal diagnostic and calibration cycle for the
oscilloscope. For the Agilent oscilloscope, this is referred to as Calibration.
This Calibration will take about 20 minutes. Perform the following steps:
1 Set up the oscilloscope with the following steps:
a Connect the keyboard, mouse, and power cord to the rear of the
oscilloscope.
b If a second monitor is being used, connect it to the VGA connector
located near the LAN port, on the rear of the oscilloscope.
c Plug in the power cord.
d Turn on the oscilloscope by pressing the power button located on the
lower left of the front panel.
e Allow the oscilloscope to warm up at least 30 minutes prior to
starting the calibration procedure in step 2 below.
2 Locate and prepare the accessories that will be required for the
internal calibration
f Locate the calibration cable.
g Locate the two Agilent precision SMA/BNC adapters.
h Attach one SMA adapter to the other end of the calibration cable -
hand tighten snugly.
i Attach the other SMA adapter to the other end of the calibration
cable - hand tighten snugly.