164 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
19 Sink Total Jitter Tests
Probing for Sink Total Jitter Tests
When performing the sink total jitter test, the DisplayPort Electrical 
Performance Compliance Test Application will prompt you to make the 
proper connections. Your DisplayPort test environment setup on the Set 
Up tab must match the physical connection. 
Figure 54 and Figure 55 show a physical connection for making 
differential and single- ended connections.
.
Infiniium Oscilloscope
1168A or 1169A 
Probe Amplifier
N5380A SMA probe head. 
Connect the plus side of 
the probe to the plus side 
of the SMA probe head 
and the negative side of 
the probe to the negative 
side of the SMA probe head.
Lane on the test fixture
SMA to SMP cables
DUT
Figure 54 Probing for Differential Tests - Sink Total Jitter Tests (Single Connection with 
DisplayPort Test Fixture)