Technical data
204 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
23 Cable Non-ISI Jitter Tests
Probing for Cable Non-ISI Jitter Tests
When performing the cable non- ISI jitter test, the DisplayPort Electrical 
Performance Compliance Test Application will prompt you to make the 
proper connections. Your DisplayPort test environment setup on the Set 
Up tab must match the physical connection. 
Figure 75 and Figure 76 show a physical connection for making 
differential and single- ended connections.
.
Infiniium Oscilloscope
1168A or 1169A 
Probe Amplifier
N5380A SMA probe head. 
Connect the plus side of the 
probe to the plus side of the 
SMA probe head and the 
negative side of the probe to 
the negative side of the SMA 
probe head.
Lane on the W2641A test fixture
SMA to SMP cables
DUT
Figure 75 Probing for Differential Tests -Cable Non-ISI Jitter Tests (Single Connection 
with W2641A DisplayPort Test Fixture)