Technical data

46 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
5 Source Non-ISI Jitter Differential Tests
Probing for Source Non-ISI Jitter Differential Tests
When performing the non- ISI jitter test, the DisplayPort Electrical
Performance Compliance Test Application will prompt you to make the
proper connections. Your DisplayPort test environment setup on the Set
Up tab must match the physical connection.
Figure 13 and Figure 14 show a physical connection for making
differential and single- ended connections.
.
Infiniium Oscilloscope
1168A or 1169A
Probe Amplifier
N5380A SMA probe head.
Connect the plus side of the
probe to the plus side of the
SMA probe head and the
negative side of the probe to
the negative side of the SMA
probe head.
Lane on the W2641A test fixture
SMA to SMP cables
DUT
Figure 13 Probing for Differential Tests - Non-ISI Jitter Tests (Single Connection with
W2641A DisplayPort Test Fixture)