Technical data

88 Agilent U7232A DisplayPort Electrical Performance Compliance Test Application
10 Source Inter Pair Skew Differential Tests
Probing for Source Inter Pair Skew Differential Tests
When performing the inter pair skew test, the DisplayPort Electrical
Performance Compliance Test Application will prompt you to make the
proper connections. Your DisplayPort test environment setup on the Set
Up tab must match the physical connection.
Figure 28 and Figure 29 below show the differential and the single- ended
connections for Inter Pair Skew Differential Tests..
Infiniium Oscilloscope
1168A or 1169A
Probe Amplifier
N5380A SMA probe head.
Connect the plus side of the
probe to the plus side of the
SMA probe head and the
negative side of the probe to
the negative side of the SMA
probe head.
Lane on the W2641A test fixture
SMA to SMP cables
DUT
Figure 28 Probing for Differential Tests - Inter Pair Skew Tests (Single Connection with
W2641A DisplayPort Test Fixture)