long channel = 1, gate =2; 
AcqrsD1_configAvgConfig(instrID, channel, "GateType", &gate); 
long preSamples = 0, postSamples = 0, maxGates = 1; 
double threshold = 0.0; // in Volts 
status = AcqrsD1_configAvgConfig(instrID, channel,  
           "PreSamples", &preSamples); 
status = AcqrsD1_configAvgConfig(instrID, channel,  
           "PostSamples", &postSamples); 
status = AcqrsD1_configAvgConfig(instrID, channel,  
           "Threshold",  &threshold); 
status = AcqrsD1_configAvgConfig(instrID, channel,  
           "NbrMaxGates", &maxGates); 
3.5.3.  SSR Time stamps 
The „On-board‟ 10 MHz reference clock is used to increment a counter. The value of the counter is stored after the 
trigger of each new segment. The value of the counter can be read by the software as shown here: 
double SSRtimeStamp; 
Acqrs_getInstrumentInfo(instrID, "SSRTimeStamp", &SSRtimeStamp); 
Since each channel is controlled by its own FPGA the time stamps for the same segment are not necessarily the same 
for the two channels. The command above works with the stamp of Channel 2.  
In this release if the function is called before the first acquisition has been started the value returned will be 0! 
It is possible to reset the time stamp when starting an acquisition using a call like 
status = AcqrsD1_acquireEx(instrID,0,4,0,0); 
or with a hardware signal on the P1 or P2 connectors. This can be done with a call like: 
Long TSReset = 1; 
status = AcqrsD1_configAvgConfig(instrID, channel, "P1Control",&TSReset); 
3.6. Configuring AP family Peak
TDC
 Analyzers 
Since Peak
TDC
 processing can be viewed as an additional form of processing after SSR acquisition please refer to the 
discussion in 3.5 Configuring SSR Analyzers for that part of process. In addition you will have to  
// Configure peak detection 
long numberOfTriggersPerSeg = 50, numberOfSegments = 1; 
double startDelta=0.1,validDelta = 0.2;  
AcqrsD1_configMode(instrID, 5, 0, 0); 
AcqrsD1_configAvgConfig(instrID, 0, "NbrRoundRobins", &numberOfTriggersPerSeg); 
AcqrsD1_configAvgConfig(instrID, 0, "NbrSegments", &numberOfSegments); 
AcqrsD1_configAvgConfig(instrID, 0, "StartDeltaPosPeakV", &startDelta); 
AcqrsD1_configAvgConfig(instrID, 0, "ValidDeltaPosPeakV", &validDelta); 
// Configure histogram 
long tdcMode = 1, tdcDepth = 1, tdcIncr = 2, tdcType = 1; 
AcqrsD1_configAvgConfig(instrID, 0, "TdcHistogramMode", &tdcMode); 
AcqrsD1_configAvgConfig(instrID, 0, "TdcHistogramDepth", &tdcDepth); 
AcqrsD1_configAvgConfig(instrID, 0, "TdcHistogramIncrement", &tdcIncr); 
AcqrsD1_configAvgConfig(instrID, 0, "TdcProcessType", &tdcType); 
// if the acquisition has segments to be histogrammed independently 
long tdcOverlay = 0; 
AcqrsD1_configAvgConfig(instrID, 0, "TdcOverlaySegments", &tdcOverlay); 