User manual
Programmer’s Guide    Page 33 of 66 
(C) Waiting for Interrupt: 
ViStatus status; 
long timeOut = 100; //   
status = AcqrsD1_waitForEndOfAcquisition(instrID, timeOut); 
if (status == ACQIRIS_ERROR_ACQ_TIMEOUT) // timeout, stop  
  STOP ACQUISITION 
This method combines low CPU usage with very good response time: 
The  function  enables  the instrument‟s 'end-of-acquisition'  interrupt  and sets  up  a semaphore  that  waits for  this 
interrupt. It then releases the thread by 'going to sleep', thus letting other threads of the application or other 
applications use the CPU time. The function returns as soon as the interrupt occurs or when the timeout expires. Note 
that the timeout asked for will be clipped to a maximum value of 10 seconds. 
The equivalent of the above for a Time-to-Digital Converter would be: 
ViStatus status; 
long timeOut = 100; //   
status = AcqrsT3_waitForEndOfAcquisition(instrumentID, timeOut); 
  if (status == ACQIRIS_ERROR_ACQ_TIMEOUT) // timeout, stop  
    status = AcqrsT3_stopAcquisition(instrumentID); 
We recommend  using  AcqrsXX_waitForEndOfAcquisition  since  it  is the  most efficient  method.  The interrupt 
latency is of the order of several  s, and no CPU time is wasted. 
3.10.4. Stopping/Forcing a D1-style Acquisition 
The  previous  section  shows  a  case  where  an  ongoing  acquisition  must  be  stopped,  typically because  there  is  no 
trigger.  Also,  in  some  situations  you  may  want  to  use  the  digitizer  to  generate  a  system  trigger  under  software 
control. 
If you still would like to have a valid snapshot of the current input signal, you should generate a trigger signal by 
software, with the function AcqrsD1_forceTrig or AcqrsD1_forceTrigEx. Typically, the acquisition does not stop 
immediately, since the digitizer may continue acquiring some additional data, depending on the delayTime and the 
data acquisition time that  were initially configured. Thus, the application should  again wait  for the acquisition to 
terminate.  Forcing  a  trigger  does  not  make  sense  for  averagers  and  analyzers  and  should  not  be  done. 
AcqrsD1_forceTrigEx allows you to  generate a trigger out signal which can be synchronized  with the sampling 
clock if desired. 
Use the following code fragment to replace STOP ACQUISITION in the previous section:  
{ 
 AcqrsD1_forceTrig(instrID); 
 if (AcqrsD1_waitForEndOfAcquisition(instrID, timeOut) == 
 ACQIRIS_ERROR_ACQ_TIMEOUT) 
 { 
 AcqrsD1_stopAcquisition(instrID); 
 SCREAM, because a major error occurred 
 } 
} 
Note that no timeout should ever occur when waiting for a 'forceTrig' to terminate, provided that the timeOut value 
was made large enough. If a timeout does occur, this would indicate a failure in the digitizer or the entire system. 
For users generating triggers under software control it may be desirable to do the data readout in a way that just gives 
the acquired data points and ignores the correction of the data gotten from the horPos measurement of the time from 
the trigger to the next data sample. This can be done using the flags parameter of the AqReadParameters structure. 