Graphics Tablet User's Manual
User Manual: Family of 8-bit Digitizers    Page 52 of 66 
Buffer size 
Upon startup of the application, the available physical memory on the computer is checked. The total data buffer size 
is then limited to the available memory minus 10% of the total memory. It is also possible to limit the total data 
buffer size to a fixed value with the -bn option (where n is the limit in MB, e.g. -b200 will limit the buffer to 200 
MB). The data buffer is then shared among the active channels. 
When the total acquisition memory (sum over all active channels) exceeds the data buffer size limit, readout of the 
data will be partial. The special operation modes such as Segments Average, Average, Digital Filter and FFT are not 
supported with partial data readout. 
Partial readout is controlled with the Very Long Memory panel (menu Options ->Very Long Memory...). The first 
segment and the first sample to read out can be specified. Note that if the buffer is large enough for at least 1 full 
segment, only full segments will be read (i.e. first sample control is ignored). 
When saving data to file, you will be prompted for a maximum number of segments per file. 
Banks for SMAR 
Some of the 10-bit Family modules are capable of acquiring data and being read out simultaneously. The -Bn option 
allows the internal memory of the digitizer to be divided into n banks. To observe the improved acquisition trigger 
rate achievable in this mode it is suggested to set up the desired conditions and then turn the event display off. The 
trigger out signal can be used to monitor the trigger rate; an interval width trigger can be used to check if triggers are 
missing. 
Calibration 
Before operation, the instruments must undergo an internal calibration. As this calibration can only proceed 
sequentially for each channel, the initialization time of AcqirisLive can become significant for systems with many 
channels. Therefore, it can sometimes be desirable to be able to start the application quickly by skipping the 
calibration (which must then be done "manually" before collecting correct data). 
DMA Data Transfers 
By default, data transfers on the PCI bus between the digitizer memory and the PC are done in DMA (Direct 
Memory Access) mode, which allows the fastest transfer speeds (up to about 130 MB/s peak, 100 MB/s sustained 
acquisition data rate). These very high data rates are not well supported by all hardware configurations (PC 
motherboard), and system hangs due to endless retries on the PCI bus have been observed on some machines. Such 
problems disappear when switching DMA mode off (data transfer rate is then about 2.8 MB/s). 
Please note that switching DMA data transfer mode off precludes the use of acquisition interrupts. The –i switch will 
therefore be ignored when –n is used. 
Process Priority 
When the program checks for an event from the operating system (e.g. to react to user input), it can put itself to sleep 
for a specified period of time. This gives other applications more processor time, but limits the performance of 
AcqirisLive in terms of maximum acquisition rate. The default is not to go to sleep, but this can be overridden to 
allow better performance for other applications. Note that this sleep policy is only in effect while the acquisition is 
running. When the acquisition stops, the sleep policy is always reset to the longest sleep period. 
Acquisition Interrupts 
When performing acquisitions, there are 2 possible ways to wait for the acquisition to complete: polling or using a 
hardware interrupt. 
•  -p Polling: AcqirisLive runs as a single thread process. After an acquisition is started, the instruments are 
regularly polled to check when the acquisition is complete. The user interface is still alive, but the latency 
for reading the data after the acquisition may vary. 
•  -i Hardware interrupt: In this mode, AcqirisLive is run as a multi-thread process. After starting an 
acquisition, a new thread is created to wait for the hardware interrupt indicating that the acquisition 
completed. This thread is put to sleep, and wakes up upon reception of the interrupt from the instrument, or 
after a 2 seconds timeout. 
Please note that switching the DMA data transfer mode off precludes the use of acquisition interrupts. The –i switch 
will therefore be ignored when –n is used. 
Transient Recorder mode 
The application will start in the Transient Recorder mode as opposed to the oscilloscope mode 