User`s guide
  115
Triggering Examples
Single-Machine Trigger Examples
Because both timers start when entering sequence level 2, they start as 
soon as the positive edge of the pulse occurs. Once the negative edge 
occurs, the sequencer transitions to level 3. If at that point, the 
MIN_WID timer is less than 496 ns or the MAX_WID timer is greater 
than 1 ms, the pulse width has been violated and the analyzer triggers. 
Otherwise, the sequence is restarted.
Measurement of Minimum and Maximum Pulse Width Limits
Triggering when a Pulse Exceeds Minimum or Maximum Limits