User`s guide
  121
Triggering Examples
Cross-Arming Trigger Examples
To look at control and status signals during 
execution of a routine
The state analyzer will trigger on the start of the routine whose control 
and status signals are to be examined more frequently than once per 
bus cycle. When the state analyzer triggers, it sends out an arm signal. 
The timing analyzer triggers when it receives the true arm level and 
detects the transition represented by Edge1. 
1 Set up one state analyzer and one timing analyzer.
2 Go to the state analyzer's Trigger menu and define term 
R_START to represent the starting address of the routine.
3 Under State Sequence Levels, enter the following sequence 
specification:
• While storing "anystate" TRIGGER on "R_START" Occurs "1" Else on "no 
state" go to level "1"
4 Go to the timing analyzer's Trigger menu.
5 Define the Edge1 term to represent a transition on one of the 
control signals.
6 Set the timing analyzer to be run by the state analyzer. Under 
Timing Sequence Levels, enter the following sequence 
specification:
•TRIGGER on "arm  Edge1" 1 time 
You do not need to use a combination trigger when one analyzer is 
armed from the other analyzer - the arm term is ANDed automatically 
with the term already in use at that level.