User`s guide
146   
Using the Oscilloscope
Using the Oscilloscope
Using the Oscilloscope
This chapter covers the oscilloscope common menus and calibration. 
This chapter covers: 
• calibrating the oscilloscope
• oscilloscope common menus.
The oscilloscope circuitry requires an operational accuracy calibration 
by the user or service department under any of the following 
conditions:
• at six month intervals or every 1000 hours
• if ambient temperature changes more than 10°C from the temperature at 
full calibration
• to optimize measurement accuracy.
To test the oscilloscope circuitry against specifications (full 
calibration), refer to chapter 3, Testing Performance in the “Agilent 
Technologies 1670G-Series Logic Analyzers Service Guide.”