User`s guide
  147
Using the Oscilloscope
Calibrating the oscilloscope
Calibrating the oscilloscope
Equipment Required
Calibration PROTECT/UNPROTECT switch
The 1670G-series logic analyzers have a calibration 
PROTECT/UNPROTECT switch on the back panel. This switch must 
be set to UNPROTECT before new calibration values from the 
operational accuracy calibration can be stored to nonvolatile RAM.
Set up the equipment
1 Turn on the logic analyzer. Let it warm up for 30 minutes if you 
have not already done so.
2 Facing the front panel, reach around to the lower right corner on 
the back of the logic analyzer. Flip the PROTECT/UNPROTECT 
switch to UNPROTECT (flip the switch up).
Equipment Critical Specification
Recommended 
Agilent Model/Part
Qty
Cable (2) BNC, 9-inch (equal length) 10502A  1
Cable 50 W BNC (m-to-m) 48-inch 10503A 1
Adapter BNC tee (m)(f)(f) 1250-0781 1
Adapter BNC (f)(f) (ug-914/u) 1250-0080 1