User`s guide
148   
Using the Oscilloscope
Calibrating the oscilloscope
Load the default calibration factors
Note that once the default calibration factors are loaded, all 
calibrations must be done. This includes all of the calibrations in the 
Self Cal menu. The calibration must be performed in the exact 
sequence listed below.
NOTE: The calibration PROTECT/UNPROTECT switch on the back of logic analyzer 
must be set to UNPROTECT.
1 Go to the Scope Calibration menu.
2 Select the Mode field, then select Service Cal from the pop-up 
menu.
3 Select the Procedure field, then select Default Values from the 
pop-up menu. 
4 Select the Start field and follow the instructions on the display. 
Loading the Default Calibration Factors
After you select the Start field, you can abort the calibration procedure 
by selecting either the Mode or Procedure fields if the Continue field is 
still displayed on the screen.