User`s guide
154   
Using the Oscilloscope
Oscilloscope Common Menus
Autoscale
Autoscale is an algorithm that automatically optimizes the display of 
one or more waveforms. When you select the Autoscale field and 
choose Continue, the autoscale algorithm starts.
What the Autoscale algorithm does when a signal is found 
The autoscale algorithm first checks all input channels to determine 
whether or not there are any signals present. The vertical scaling is 
then set as required for each channel. Next the time base is scaled for a 
single input channel, the trigger channel is selected, the data is 
acquired, and the waveforms are displayed.
Finding the vertical settings. The autoscale algorithm sets the 
vertical scaling (V/Div and offset) appropriate to the input signal for 
channel 1. This process is repeated for channel 2.
Finding the time base settings. The time base settings (s/Div and 
delay) are determined based on the input signal of Channel 1 (if active) 
or Channel 2. The time base is scaled so that between two and five 
complete cycles of the time base scaling source input signal can be 
seen on the screen. The trigger settings are also changed by the 
autoscale algorithm and an edge mode trigger with CHAN 1 or CHAN 2 
is selected. The channel selected for time base scaling is selected as 
the trigger source.