Using the Oscilloscope
The Scope Trigger Menu
Auto-Trig field
The Auto-Trig field allows you to specify whether or not the 
acquisitions should wait for the specified trigger condition to occur. 
When you select the Auto-Trig field, the field toggles between On and 
Off. The On and Off fields are discussed below.
The default selection for the Auto-Trig field is On.
On
When you set auto-trigger to On, the oscilloscope waits 50 ms (20-Hz 
rate) for a trigger to occur. If a trigger does not occur within that time, 
the current contents of acquisition memory are displayed. The 
message "Auto triggered" is displayed if one of the following conditions 
occurs: 
• No signal is on the input. In this case, the oscilloscope will display a 
baseline. 
• There is a signal but the specified trigger condition has not been met 
within 50 ms. In this case, the waveform display will not be synchronized 
to a trigger point.
Off
When you set auto-trigger to Off, the oscilloscope waits until a trigger 
is received before the waveform display is updated. If a trigger does not 
occur, the screen is not updated and the message "Waiting for Trigger" 
is displayed. Use this mode when: 
• The trigger source signal has less than a 20-Hz repetition rate.
• The trigger events counter (refer to "Count field") is set so that the 
number of trigger events would not occur before 50 ms.
• When you want to trigger on a specific event only.