User`s guide
  259
Logic Analyzer Reference
Probing
Oscilloscope probes (oscilloscope option only)
The two oscilloscope probes supplied with the oscilloscope option are 
Agilent Technologies 1160A Miniature Passive Probes. These small, 
lightweight probes allow measurements that were previously very 
difficult in densely populated circuits.
For complete information on the operation, maintenance, and 
adjustments of the miniature passive probes, be sure to read the 
operating note that is packaged with the probes.
Probe Inputs
Probe inputs are located on the front panel to the right of the power 
switch. Input 1 is on the left. The probes may be connected directly to 
the BNC input connectors. The signal is dc-coupled to the oscilloscope.
BNC cables can be connected directly to the BNC connectors. A BNC-
to BNC cable is not provided with the instrument, but you can order it 
separately.
Maximum Probe Input Voltage
The maximum input voltage of each logic analyzer probe is ±40 volts 
peak. The maximum input voltage of the oscilloscope is ±250 volts dc 
at 1 MΩ setting and 5 volts rms at 50Ω setting.
Calibration Outputs
There is one calibration output BNC located on the rear panel. It is the 
AC/DC calibration signal source. This signal is used during calibration 
of the oscilloscope. This calibration signal can also be used for probe 
compensation adjustment.