User`s guide
298   
Logic Analyzer Reference
The Analyzer Format Menu
State acquisition modes 
The State Acquisition Mode field identifies the channel width and 
memory depth of the selected acquisition mode. There are two 
configurations of channel width/memory depth. 
Full Channel/64K Memory/150 MHz
Full-channel mode uses both pods in a pod pair for 34 channels of 
width and a total memory depth of 64 K per channel. If time or state 
tags are turned on, the total memory is evenly split between data 
acquisition storage and time or state tag storage. To maintain the full 
64 K per channel depth with state or time tags, leave one pod pair 
unassigned.
Half Channel/128K Memory/150 MHz
Half-channel mode cuts the channel width to 17 channels. In this 
mode, the pod used within the pod pair is selected through the Pod 
field. In half-channel mode, the memory depth is increased to 128 K 
per channel. Time or state tags are not available in this mode.