User`s guide
312   
Logic Analyzer Reference
The Analyzer Trigger Menu
The Analyzer Trigger Menu
Trigger sequence levels
Sequence levels are the definable stages of the total trigger 
specification. Individual sequence levels are assigned using either a 
predefined trigger function or a user-level trigger function. The total 
trigger specification can contain both kinds of function.
See Also "Using the Trigger Menu" on page 70 for more on setting up a trigger.
Sequence level usage
Generally, you would think using one function in one sequence level 
uses up one of the available sequence levels. This may not always be 
the case. Some of the more complex predefined functions require 
multiple sequence levels. Keep this point in mind if you are near the 
limit on remaining sequence levels. The exact number of internal levels 
required per function, and the remaining available levels, are shown 
within the function library list. User functions, however, use only one 
level almost all the time. The only instance where multiple levels are 
used with the User function is when the "<" duration is assigned.