User`s guide
  325
Logic Analyzer Reference
The Analyzer Trigger Menu
Using Preset Values
Assign. Assign toggles which machine the term is assigned to. All of 
the available resource terms except the Edge terms can be assigned to 
any analyzer. However, a term can only be assigned to one analyzer at a 
time.
Rename. Rename lets you change the term name. This function works 
for all terms.
Clear (=X). Clear sets the terms to their broadest possible meaning. 
For terms a - j, the assignment field is set to all Xs (don't cares). For 
Range 1 and 2, the boundaries are set to the maximum and minimum 
values. For Timers 1 and 2, the assignment field is set to a minimum 
time of 400 ns. For Edge 1 and 2, the assignment field is set to don't 
cares.
Set (=1). In terms a - j, the assignment field is set to its maximum 
value, with all bits set to 1. This option is not available for the Range, 
Timer, and Edge terms.
Reset (=0). In terms a - j, the assignment field is set to its minimum 
value, with all bits set to 0. This option is not available for the Range, 
Timer, and Edge terms.
Assigning Bit by Bit
Bit pattern terms. Just to the right of the bit pattern name fields are 
the term assignment fields. When any of the individual assignment 
fields are selected, a keypad appears. Use this keypad to assign real 
values or Don't Care (X) values.
Edge terms. Assign edge terms the same way you do bit pattern 
terms. Edge terms can be used singularly or in combination with each 
other across all assigned channels. When you specify an edge on more 
than one channel, the analyzer ORs the edges.