User`s guide
330   
Logic Analyzer Reference
The Analyzer Trigger Menu
In a timing analyzer, even when the trigger position is set to Start or 
End, there will always be a small portion of pre-trigger and post-trigger 
data stored. Most of the choices designate prestore and poststore 
percentages, but the Delay setting affects when the memory begins 
storing data relative to the trigger.
Delay (Timing only)
The Delay field delays the start of acquisition storage after the trigger. 
The delay time range is affected by the sample period but could range 
from 16 ns to 8 ks. As the picture in the pop-up menu shows, any data 
falling between the trigger and the delay time is not stored.
Branches Taken Stored / Not Stored field
The Branches Taken field is a toggle field that sets the analyzer to 
store, or not to store, the resource term that sent the analyzer off on a 
branch.
As the analyzer steps through the sequence instructions, it may take 
either branch of a sequence level. With Branches Taken set to Stored, 
the state data values that caused the branch are stored. When the 
analyzer is set to Branches Taken Not Stored, only the data you 
explicitly designate in the sequence levels is stored. 
The Branches Taken Stored/Not Stored is not available when the 
analyzer is configured as a timing analyzer. 