User`s guide
334   
Logic Analyzer Reference
The Waveform Menu
The Waveform Menu
sec/Div field
When acquisition control is set to automatic, the sec/Div field affects 
the sample period. Timing waveforms are reconstructed relative to the 
sample period. A shorter sample period puts more sample points on the 
waveform for a more accurate reconstruction but also fills memory 
more quickly.
If the sec/Div is changed resulting in a change in the next sample 
period, you must run the analyzer again before the current sample 
period display is updated. The sample period is shown in the second 
row from the top when the markers are turned off.
Accumulate field
The Accumulate field controls whether old data is cleared or displayed 
along with new data. The Accumulate field will toggle On/Off. When 
Accumulate is on, the analyzer displays the data from a current 
acquisition on top of the previously acquired data. When Accumulate is 
off, the display is cleared before each new run cycle.
If you leave the Waveform menu, or pop up a menu over the waveform 
display, any accumulated display data is lost and the accumulation 
process starts over.