Logic Analyzer Reference
The Waveform Menu
Delay field
Depending on the analyzer configuration, a positive or negative delay 
measured in either states (State only) or time (Timing only) can be set. 
The Delay field lets you scroll the data and place the display window at 
center screen. Changing the delay will not affect the data acquisition 
unless it is a timing analyzer and the acquisition mode is automatic. In 
this case, the sample period changes.
The delay range of a timing analyzer is from -2500 seconds to +2500 
seconds. The delay range of a state analyzer is dependent on memory 
length and cannot exceed total memory size.
Waveform label field
The waveform label field, located on the left side of the waveform 
display, is both a display and configuration field. After all desired 
waveforms are configured for display, they are listed in the waveform 
label field. If there are more waveforms than can be displayed, you can 
roll the list by selecting the waveform label field, then after the roll 
indicator appears, turn the knob.
If the waveform label field is selected a second time, a waveform 
modification menu appears. Use this menu to configure the waveform 
display.
When the waveform modification menu appears, select the operation 
to insert, replace, delete, or scale waveforms into the display. You can 
display a maximum of 24 waveforms on screen at one time. 