System Performance Analysis (SPA) Software
System Performance Analysis Software
SPA Time Interval Menu 
For measurement purposes, the analyzer measures the time between 
the first occurrence of the Start condition and the first occurrence of 
the Stop condition.
Defining the Time Interval ranges. Before changing the ranges 
from their default values, you may want to press Run and acquire some 
data. From this initial run, the Maximum (Max), Minimum (Min), and 
Average (Avg) statistics on the display will help you choose the 
appropriate set of Time Interval ranges.
To define the ranges, select the fields for the lower and upper limits 
and enter the limits of the range. The ranges do not need to be 
contiguous, but if you leave gaps between the ranges, critical data may 
be missed. Also, if you overlap ranges, data may be counted multiple 
times and present a misleading histogram.