User`s guide
  427
Troubleshooting the Logic Analyzer
Analysis Probe Problems
Slow clock
If you have the analysis probe hooked up and running and observe a 
slow clock or no activity from the interface board, the +5 V supply 
coming from the analyzer may not be getting to the interface board.
❏ To check the +5 V supply coming from the analyzer, disconnect one of the 
logic analyzer cables from the analysis probe and measure across pins 1 
and 2 or pins 39 and 40.
• If +5 V is not present, check the internal analysis probe fuse or current 
limiting circuit on the logic analyzer. For information on checking this fuse 
or circuit, refer to the 1670G-Series Logic Analyzers Service Guide.
• If +5 V is present and the cable connection to the analysis probe appears 
sound, contact your nearest Agilent Technologies Sales Office for 
information on servicing the board.