Troubleshooting the Logic Analyzer
Error Messages
Error Messages
This section lists some of the messages that the analyzer displays when 
it encounters a problem.
". . . Inverse Assembler Not Found"
This error occurs if you rename or delete the inverse assembler file 
that is attached to the configuration file. Ensure that the inverse 
assembler file is not renamed or deleted, and that it is on the same 
flexible disk or in the same directory as the configuration file.
"No Configuration File Loaded"
This is usually caused by trying to load a configuration file for one type 
of module or the system into a different type of module.
Verify that the appropriate module has been selected as the target of 
the Load operation. Selecting Load All will cause incorrect operation 
when loading most analysis probe configuration files.
See Also "To Load a Configuration" on page 137.