User`s guide
  455
Operator’s Service
Preparing For Use
Preparing For Use
This section gives you instructions for preparing the logic analyzer for 
use. 
Power requirements
The logic analyzer requires a power source of either 115 VAC or 230 
VAC, -22 % to +10%, single phase, 48 to 66 Hz, 200 Watts maximum 
power. 
Category I: Signal level, telecommunications, electronic. (Seperated 
from line voltage by transformer, etc).
Category II: Local level, appliances, portable equipment (line voltage in 
appliance and to wall outlet).
Operating environment
The operating environment is listed in previous chapter. Note the 
noncondensing humidity limitation. Condensation within the 
instrument can cause poor operation or malfunction. Provide 
protection against internal condensation.
The logic analyzer will operate at all specifications within the 
temperature and humidity range given. However, reliability is 
enhanced when operating the logic analyzer within the following 
ranges:
• Temperature: +20°C to +35 °C (+68 °F to +95 °F)
• Humidity: 20% to 80% noncondensing
• Pollution degree 2: Normally only dry non-conductive pollution occurs. 
Occasionally a temporary conductivity caused by condensation must be 
expected.