User`s guide
  81
Using the Trigger Menu
Setting Up Time Correlation between Analyzers
Setting Up Time Correlation between Analyzers
There are two possible combinations of analyzers: state and state, and 
state and timing. Timing and timing is not possible because the 
Analyzer Configuration menu only permits one analyzer at a time to be 
configured as a timing analyzer. For either combination, time 
correlation is necessary for interleaving and mixed display.
Time correlation is useful when you want to store different sorts of 
data for each trace, but see how they are related. For instance, you 
could set up a timing and a correlated state analyzer and see if setup 
and hold times are being met. Or, you could set up two state analyzers 
and have one watch normal program execution and the other watch 
the control and status lines.
Time correlation requires that state analyzers store time tags. You set 
the state analyzer to store time tags by turning on Count Time in the 
Analyzer Trigger menu. The timing analyzer already stores time tags 
when it samples data.
See Also "Special displays" on page 128 for more information on interleaving and 
mixed display.