User`s guide
Contents ▲ 67 ▼ Index
Essential Measurement Practices (Electrophoretic Assays)
General:
WARNING
Wear hand and eye protection and follow good laboratory practices when preparing 
and handling reagents and samples.
WARNING
No data is available addressing the mutagenicity or toxicity of the dye/DMSO 
reagent. Because the dye binds to nucleic acids, it should be treated as a potential 
mutagen and used with appropriate care. The DMSO mixtures should be handled 
with particular caution as DMSO is known to facilitate the entry of organic 
molecules into tissues. We strongly recommend using double gloves when handling 
DMSO mixtures.
• Handle and store all reagents according to the instructions given in the Reagent Kit 
Guides.
• Avoid sources of dust or other contaminants. Foreign matter in reagents and samples 
or in the wells of the chip will interfere with assay results.
• Always insert the pipette tip to the bottom of the well when dispensing the liquid. 
Placing the pipette at the edge of the well may lead to poor results due to the formation 
of a bubble on the bottom of the well.