User`s manual

28
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ info@datatec.de ▪ www.datatec.de
28
Waveform measurements and cursors
DC vertical accuracy/cursors** Single cursor accuracy: ± [DC vertical gain accuracy + DC vertical offset accuracy +
0.21% full scale]
Dual cursor accuracy: + [DC vertical gain accuracy + 0.42% full scale]*
Cursors 2 pairs of XY cursors
Automatic measurement of positions, ∆X, 1/∆X, ∆Y, and ∆Y/∆X
Automatic measurements Measurements continuously updated with statistics. Cursors track last selected
measurement. Select up to four measurements from the list below:
Voltage Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot,
average- N cycles, average- full screen, DC RMS- N cycles, DC RMS- full screen,
AC RMS- N cycles, AC RMS- full screen (standard deviation), ratio (RMS1/
RMS2).
Time Period, frequency, counter, + width, - width, burst width, duty cycle, rise time,
fall time, delay, phase, X at min Y, X at max Y
Count Positive pulse count, negative pulse count, rising edge count, falling edge count
Mixed Area- N cycles, area- full screen
Counter Built-in frequency counter
Source Any analog or digital channel
Resolution 5 digits. Up to 8 digits with an external reference 10 MHz input.
Max frequency Bandwidth of oscilloscope
Mask limit test option Standard mask/limit test capability that provides easy, fast pass/fail comparison
of a signal under test to a predefined mask template or auto-mask template.
Predefined mask templates or edits to an auto-mask template can be made via a
simple text editor. > 270,000 mask tests per second (waveform update rate)
Oscilloscope Experience Redefined:
Waveform math
Number of math functions Four (display one at a time)
Arithmetic Add, subtract, multiply, FFT, differentiate, integrate, square root, divide, Ax + B, square,
absolute, common log, natural log, exponential, base 10 exponential, LP filter, HP filter,
averaged value, magnify, measurement trend, chart logic bus (timing or state)
FFT Up to 64 kpts resolution. FFT window types: hanning, flat top, rectangular, Blackman-Harris
InfiniiVision 4000 X-Series Performance Characteristics
Trigger type selections
MIL-STD1553 (Option) Trigger on MIL-STD 1553 signals on data word start/stop, command/status start/stop, RTA, RTA
+ 11 bits, and error conditions (parity, sync, Manchester)
SPI (Option) Trigger on SPI (serial peripheral interface) data pattern during a specific framing period. Supports
positive and negative chip select framing as well as clock Idle framing and user-specified number
of bits per frame. Supports MOSI and MISO data
UART/RS232/422/485 (Option) Trigger on Rx or Tx start bit, stop bit, data content, or parity error
USB (Option) Trigger on start of packet (SOP), end of packet (EOP), suspend***, resume***, reset***, packets
(token, data, handshake, or special), and errors (PID, CRC5, CRC16, glitch, bit stuff***, SE1***).
Supports USB 2.0 low speed, full speed, and hi-speed (hi-speed is supported on 1 GHz and 1.5 GHz
models only).
* Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
** 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity setting
*** Suspend, resume, reset, bit stuff error, and SE1 error are USB 2.0 low and full speed only.