User`s manual

29
dataTec ▪ Ferdinand-Lassalle-Str. 52 ▪ 72770 Reutlingen ▪ Tel. 07121 / 51 50 50 ▪ Fax 07121 / 51 50 10 ▪ info@datatec.de ▪ www.datatec.de
29
Display characteristics
Display 12.1-inch high-definition capacitive touch display
Resolution 800 (H) x 600 (V) pixel format (screen area)
Graticules 8 vertical divisions by 10 horizontal divisions with intensity controls.
Format YT, XY and Roll
Maximum waveform update rate > 1,000,000 wfm/s
Persistence Off, infinite, variable persistence (100 ms - 60 s)
Intensity gradation 64 intensity levels
Oscilloscope Experience Redefined:
Connectivity
USB 2.0 hi-speed host port USB 2.0 hi-speed host ports x3, two front and one real panel. Supports memory devices and printers.
USB 2.0 hi-speed device port
One USB 2.0 hi-speed device port on rear panel. USB Test & Measurement Class (USBTMC) compatible.
LAN port 10/100Base-T port on rear panel. LXI IPv6 extended function
Web remote control VNC Web interface (via major Web browsers)
Video out port SVGA out on rear panel. Connect oscilloscope display to an external monitor or projector.
GPIB port N4865A GPIB-to-LAN adapter (optional)
10 MHz out/in BNC connector on the rear panel. Supported modes: Off, 10 MHz out, or reference signal mode (10 MHz in).
Trigger out BNC connector on the rear panel. Supported modes: triggers, mask, waveform generator 1 sync
pulse, and waveform generator 2 sync pulse.
Dual-channel WaveGen built-in function/arbitary waveform generator (specifications are typical)
WaveGen outputs Two (front-panel BNC connectors)
Both waveform generator outputs can be frequency tracked, amplitude tracked, or completely
tracked.* A generator’s output can be inverted to create a differential signal
Waveforms Sine, square, ramp, pulse, DC, noise, sine cardinal (sinc), exponential rise, exponential fall, cardiac,
Gaussian pulse, and arbitrary
Modulation Modulation is available on channel 1 only. Modulation is not available when tracking mode is
enabled.
Modulation Types: AM, FM, FSK
Carrier Waveforms: Sine, Ramp, Sine Cardinal, Exponential Rise, Exponential Fall, and Cardiac.
Modulation Source: Internal (no external modulation capability)
AM:
Modulation: Sine, Square, Ramp
Modulation Frequency: 1Hz to 20 kHz
Depth: 0% to 100%
FM:
Modulation: Sine, Square, Ramp
Modulation Frequency: 1Hz to 20 kHz
Minimum Carrier Frequency: 10Hz
Deviation: 1 Hz to Carrier Frequency or (2e
12
/ Carrier Frequency), whichever is smaller
FSK:
Modulation: 50% duty cycle square wave
FSK Rate: 1Hz to 20 kHz|
Hop Frequency: 2 x FSK Rate to 10 MHz
InfiniiVision 4000 X-Series Performance Characteristics
* Only the following combination of wave shapes can be frequency tracked or completely tracked: 1) sine, ramp, sine cardinal, cardiac, and Gaussian
pulse. 2) square wave and pulse. 3) exponential rise and exponetial fall. 4) arbitrary.