ABORt
The ABORt command subsystem removes the multimeter from the wait-for-trigger
state and places it in the idle state.
ABORt can only
be used with the following trigger sources:
TRIGger:SOURce BUS
and
TRIGger:SOURce HOLD.
Subsystem Syntax ABORt
Comments
ABORt
does not affect any other settings of the trigger system. When the INITiate
command is sent, the trigger system will respond as it did before ABORt was
executed.
When TRIGger:SOURce BUS is selected, ABORt returns the multimeter to the
idle state. When a Group Execute Trigger (
GET) bus command or *TRG common
command is executed, the
"Trigger ignored" error is generated.
When TRIGger:SOURce HOLD is selected, ABORt returns the multimeter to the
idle state. All subsequent single triggers sent using
TRIGger:IMMediate are
ignored and the "Trigger ignored" error is generated.
When the trigger system is initiated from the HP-IB interface, execute the HP-IB
CLEAR command or press the HP E1301A front panel “Clear Instr” or
Reset Instr” key to return to the idle state.
When the trigger system is initiated from the HP E1301A front panel, execute
*RST over the HP-IB interface or press the HP E1301A front panel “Clear Instr
or “Reset Instr” key to return to the idle state.
Related Commands: INITiate, TRIGger:SOURce
*RST Condition: After a *RST, the multimeter acts as though an ABORt has
occurred.
Example Aborting a Measurement
CONF:VOLT:DC (@100:104) !Function: DC voltage on specified
channels.
TRIG:SOUR HOLD !Suspend triggering; wait for TRIG:IMM
command.
INIT !Place multimeter in wait-for-trigger state.
ABOR !Place multimeter in the idle state.
Chapter 5 Command Reference 122