User`s manual

CALibration
The CALibration command subsystem selects the multimeter’s line reference
frequency (
CALibration:LFRequency) and enables/disables the autozero mode
(
CALibration:ZERO:AUTO).
Subsystem Syntax CALibration
:LFRequency <frequency>
:LFRequency? [MIN | MAX]
:ZERO:AUTO <mode>
:ZERO:AUTO?
:LFRequency
CALibration:LFRequency <frequency> selects the line reference frequency used
by the multimeter’s analog-to-digital converter.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<frequency>
numeric 50 | 60 | MIN | MAX hertz
Comments
MIN
selects the minimum line reference frequency (50 Hz). MAX selects the
maximum line reference frequency (60 Hz).
The line reference frequency is set to 60 Hz at the factory. The setting is stored in
non-volatile mainframe memory and is changed only when
CAL:LFR is executed.
For a line frequency of 400 Hz, the 50 Hz reference frequency is used; however,
since 50 Hz is a subharmonic of 400 Hz, it provides normal mode rejection of
power line related noise.
*RST Condition: The selected line reference frequency remains unchanged
since it is stored in non-volatile mainframe memory.
Example Selecting the Line Frequency Reference
CAL:LFR 50 !Reference frequency is 50 Hz.
:LFRequency?
CALibration:LFRequency? [MIN | MAX] returns one of the following numbers to
the output buffer:
The present line reference frequency ("50" or "60") if MINimum or MAXimum is
not specified.
The minimum line reference frequency ("50") if MIN is specified.
The maximum line reference frequency ("60") if MAX is specified.
123 Command Reference Chapter 5