Related Commands: FETCh?, INITiate, READ?
Example Making 4-Wire Ohms Measurements
CONF:FRES 1560,MAX,(@100:103) !Function: 4-wire ohms;
range selected: 1861
;
MAX resolution: 125 m
;
specify sense channel list.
TRIG:COUN 3 !Scan channel list 3 times (take 4 readings
per trigger); trigger source is IMMediate by
default.
READ? !Place multimeter in wait-for-trigger state and
make measurements;
send readings to output buffer.
enter statement !Enter readings into computer.
:RESistance
CONFigure:RESistance [<range>[,<resolution>]] , <channel_list>
selects the 2-wire ohms function and allows you to specify the range and resolution.
Two-wire resistance measurements can only be made using the scanning multimeter
(a channel list is required).
For a complete listing of range and resolution values available, see
Table 5-1 on page 120.
Parameters
Parameter Name Parameter Type Range of Values Default Units
<range>
numeric
232 | 1861 | 14894 |
119156 | 1048576 |
AUTO | DEF | MIN | MAX
ohms
<resolution>
numeric resolution (see Table 5-1) |
DEF | MIN | MAX
ohms
<channel_list>
numeric Refer to the multiplexer user’s manual for
a list of channels available.
none
Comments To select a standard measurement range, specify range as the input signal’s
maximum expected resistance. The multimeter then selects the correct range.
The AUTO and DEFault options for the range parameter have the same effect
(enable autorange). The
DEF option for the resolution parameter defaults the
integration time to 1 PLC.
The MIN and MAX parameters select the minimum or maximum values for range
and resolution:
For range:
MIN = 232 ; MAX = 1048576
For resolution:
MIN selects the best resolution (the smallest value from Table 5-1) for
the selected range.
MAX selects the worst resolution (the largest value from Table 5-1)
for the selected range.
Chapter 5 Command Reference 128